Ultrasonic Generator Specifications:
Frequency: 1.0 MHz 5%
3.2 MHz 5%
Modes: Continuous
Pulsed—20% duty cycle
Pulse Repetition Rate: 100 Hz 20%
Pulse Duration: 2 msec 20%, 20% duty cycle
Temporal Peak/ average
intensity ratio:
5:1 20%, 20% duty cycle
Maximum output power: 11 w with a 5 cm applicator, (me 7513)
Maximum intensity: 2.2 w/cm
Indication accuracy: 20% (for any level above 10% of maximum)
Output description: The output waveform is continuous or pulsed as
programmed by the membrane panel control. In the pulse
mode the 1 or 3.2 MHz sine wave pulses
are modulated. The power level is adjusted by varying the
pulse amplitude. The pulse waveforms are shown
below:
1 or 3.2 MHz
Ultrasound
2 ms
Pulse
Width
10 ms
Pulse Space
Output
Time
Pulse Waveform—20% Duty Cycle
In the continuous mode, the power is on at least 95% of the
time the timer is running. The continuous mode waveform
is shown below:
1 or 3.2 MHz
Ultrasound
Output
Time
Continuous Waveform
Ultrasonic Applicator Specifications:
Piezoelectric discs: The output transducer utilizes a barium titanate disc with
a specially coated face.
Individual Applicator Specifications:
Applicator Part Number Frequency Effective Radiating Area
Me 7513 1 or 3.2 MHz 5% 5 cm 20%
Maximum Beam
Non-Uniformity Ratio: 6:1
Spatial Pattern: The applicator produces a collimated (cylindrical) beam
with an area of 5 cm , measured 5 mm from the ceramic
disc surface when the radiation is emitted into the
equivalent of an infinite medium of distilled water at 30 c.
The beam of the applicator is circular in all planes parallel
to the applicator face. a few inches from the face, it is a
single smooth bell-shaped curve. Nearer the face the
pattern varies more due to phase cancellations. Sample
curves measured in the far field from the surface are shown
in Figures 9.3 and 9.4.
5 cm Applicator (1 MHz), me 7513,—Three Dimensional Beam Pattern
5 cm Applicator (3.2 MHz), me 7513,—Three Dimensional Beam Pattern
Waveform Specifications:
Interferential Mode
Interferential Waveform
Waveform Type: Sinewave
Polarity: None
Volts: 0-65 volts rms, 1 Kohm load
Current: 0-65 mA rms, 1 Kohm load
Average current at
Premodulated Mode
Premodulated Waveform
Waveform Type: Amplitude modulated
sine wave
Polarity: None
Volts: 0-50 volts rms, 1 Kohm load
Current: 0-50 mA rms,
1 Kohm load
Average current at
maximum intensity
and frequency: 50 mA rms
Maximum current
ID producto: 686092 | Referencia: 686092 | Vendido por: Ancor Comercio Internacional